| Measurement technology | light source: 3 LEDs (375 nm/470 nm/590 nm)detector: 4 photo diodes with filter | |
| Measurement principle | Fluorescence | |
| Parameter | Chlorophyll a [µg/L]Phyocyanin [µg/L]CDOM [µg/L] | |
| Measuring range | 0…200 µg/L0…200 ppb | |
| Measurement accuracy | 5 % | |
| Turbidity compensation | No | |
| Data logger | ~ 10 MB | |
| T100 response time | min. 12 s | |
| Measurement interval | min. 6 s | |
| Housing material | Stainless steel (1.4571/1.4404) or titanium (3.7035) | |
| Dimensions (L x Ø) | 155 mm x 36 mm | ~ 6.1˝ x 1.4 |
| Weight | stainless steel ~ 0.6 kgtitanium ~ 0.5 kg | ~ 1.3 lbs~ 1.1 lbs |
| Interface | digital:Ethernet (TCP/IP)RS-232 oder RS-485 (Modbus RTU, OGC PUCK) | |
| Power consumption | ≤ 1.8 W | |
| Power supply | 12…24 VDC (± 10 %) | |
| Maintenance effort | ≤ 0.5 h/month (typical) | |
| Calibration/maintenance interval | 24 months | |
| System compatibility | Modbus RTU, OGC PUCK | |
| Warranty | 1 year (EU: 2 years)US: 2 years | |
| Installation | ||
| Max. pressure | with Subconn 30 barwith fixed cable 3 barin FlowCell 1 bar, 2…4 L/min | ~ 435 psig~ 43.5 psig~ 14.5 psig, 0.5 to 1 gpm |
| Protection type | IP68NEMA 6P | |
| Sample temperature | +2…+40 °C~ +36 °F to +104 °F | |
| Ambient temperature | +2…+40 °C~ +36 °F to +104 °F | |
| Storage temperature | -20…+80 °C~ -4 °F to +176 °F | |
| Inflow velocity | 0.1…5 m/s~ 0.33 fps to 16.4 fps | |
Our high-end matrixFlu VIS fluorometer combines multiple excitation and detection wavelengths for fluorescence meas urements in a single device with a highly compact design. The special optical arrangement of excitation and detection channels enables not only single values to be determined, but also a 4×4 matrix of wavelength combinations. This allows quasi synchronous in-situ detection of EEMs (Excitation Emission Matrices).
MatrixFlu VIS is primarily designed for the on-line detection of algae (cyanobacteria, green algae, etc.) and is expanded by the detection of CDOM.State-of-the-art, specially selected LEDs are used for fluorescence excitation. The stability of measured values is increased by an internal temperature correction.
Equipped with our innovative G2 interface with web browser configuration, internal data logger, flexible protocols and data outputs, matrixFlu offers extensive features that go significantly beyond what’s available on the market today.
The unified platform of all TriOS photometers also facilitates a standardized spare parts and consumables system, which allows the use of a wide range of accessories for our devices. Furthermore the cutting-edge G2 interface enables quick integration into thirdparty systems.
“HailSens represents the culmination of years of dedicated research. As the only hail sensor that seamlessly integrates with hailpad data – the gold standard for scientists worldwide.”
“HyQuant Edge is the first radar sensor-based monitoring station for hydrology built on this architecture. Add a SIM card and power, and go from unboxing to live data in minutes.”
“Both Classic and Edge sensors have their place in modern monitoring. The goal isn’t to eliminate choice — it’s to expand it.”
Fill out the form and our experts will get back to you shortly.

Countrys
Organizations
Certified
You are currently viewing a placeholder content from YouTube. To access the actual content, click the button below. Please note that doing so will share data with third-party providers.
More InformationYou need to load content from reCAPTCHA to submit the form. Please note that doing so will share data with third-party providers.
More InformationYou need to load content from reCAPTCHA to submit the form. Please note that doing so will share data with third-party providers.
More InformationYou need to load content from reCAPTCHA to submit the form. Please note that doing so will share data with third-party providers.
More Information