| Measurement principle | Luminescence | |
| Parameter | Dissolved Oxygen | |
| Measurement range | 0…20 mg/L 0…20 ppm 0…200 % | |
| Measurement accuracy | ± 0.1 mg/L ± 0.1 ppm ± 1 % | |
| Resolution | 0.01 | |
| Measurement interval | > 5 s | |
| Resolution (temperature) | 0.01 °C | |
| Accuracy (temperature) | 0.5 °C | |
| Membrane Cap | No cross-sensitivity with: pH 1 – 14; CO2, H2S, SO2 Cross-sensitivity to Organic solvents, such as acetone, toluene, chloroform or methylene chloride Chlorine gas | |
| Material | Standard Version in passivated Stainless steel 316L body, strainer and screw, For Seawater application Version in Titanium body, strainer and screw, Cable: polyurethane jacket Steam gland: Polyamide Patch with active material (black) – Membrane: Optical isolation silicon | |
| Dimensions (L x Ø) | 146 mm x 25 mm | ~ 5.7˝ x 1˝ |
| Weight stainless steel | ~ 450 g | ~ 1 lbs |
| Weight titanium | ~ 300 g | ~ 0.7 lbs |
| Interface | RS-485 (Modbus RTU) | |
| Power consumption | 1 W | |
| Power supply | 12 V (± 10 %) | |
| Sensor cable | 2 m and 10 m | |
| Calibration/maintenance interval | 2 years | |
| Warranty | 1 Year (EU & US: 2 years) on electronics; All wearing parts are not included in the warranty | |
| Max. pressure | 5 bar | ~ 72.5 psig |
| Protection type | IP68 | NEMA 6P |
| Sample temperature | 0…+50 °C | ~ +32 °F… +122 °F |
| Ambient temperature | 0…+50 °C | ~ +32 °F… +122 °F |
| Storage temperature | -10…+60 °C | ~ +14 °F… +140 °F |
The oxygen sensor uses luminescence-based optical measurement technology and measures reliably and precisely. The low maintenance and small amount of consumable materials needed by the sensor provide immediate returns on investment. Only the membrane cap must be replaced every two years. The sensor can also be used in applications with a very weak water flow. The oxygen sensor is available with a 10-meter or a 2-meter cable.
“HailSens represents the culmination of years of dedicated research. As the only hail sensor that seamlessly integrates with hailpad data – the gold standard for scientists worldwide.”
“HyQuant Edge is the first radar sensor-based monitoring station for hydrology built on this architecture. Add a SIM card and power, and go from unboxing to live data in minutes.”
“Both Classic and Edge sensors have their place in modern monitoring. The goal isn’t to eliminate choice — it’s to expand it.”
You are currently viewing a placeholder content from YouTube. To access the actual content, click the button below. Please note that doing so will share data with third-party providers.
More InformationYou need to load content from reCAPTCHA to submit the form. Please note that doing so will share data with third-party providers.
More InformationYou need to load content from reCAPTCHA to submit the form. Please note that doing so will share data with third-party providers.
More InformationYou need to load content from reCAPTCHA to submit the form. Please note that doing so will share data with third-party providers.
More Information